An exploratory study of the impact of CT slice thickness and inter-rater variability on anatomical accuracy of malunited distal radius models and surgical guides for corrective osteotomy.
Ontology highlight
ABSTRACT: High-resolution CT images are essential in clinical practice to accurately replicate patient anatomy for 3D virtual surgical planning and designing patient-specific surgical guides. These technologies are commonly used in corrective osteotomy of the distal radius. This study evaluated how the virtual radius models and the surgical guides' surface that is in contact with the bone vary between experienced raters. Further, the discrepancies from the reference radius of surgical guides and radius models created from CT images with slice thicknesses larger than the reference standard of 0.625mm were assessed. Maximum overlap with radius model was measured for guides, and absolute average distance error was measured for radius models. The agreement between the lower-resolution guides surface and
SUBMITTER: Gryska E
PROVIDER: S-EPMC11476685 | biostudies-literature | 2024