Comparative Analysis of NiTi Instruments with Different Alloy Treatments.
Ontology highlight
ABSTRACT: This study aims to compare the cyclic fatigue resistance of nickel-titanium (NiTi) endodontic instruments, focusing on the impact of various alloy treatments and manufacturing processes across different generations of these instruments; Twenty instrumentation systems from different generations, comprising both continuous and reciprocating motion designs, were tested. Four hundred instruments underwent cyclic fatigue testing using an INSTRON machine, with the time and number of cycles to fracture (NCF) recorded for each instrument. Statistical analyses were performed to compare the fatigue resistance between systems, generations, and motion types; Instruments treated with advanced thermal processing, such as Excalibur, Reciproc Blue, and TruNatomy, demonstrated superior resistance to fractu
SUBMITTER: Aranguren J
PROVIDER: S-EPMC11477650 | biostudies-literature | 2024 Sep
REPOSITORIES: biostudies-literature