Achieving automated and high-precision in situ analysis of the dimensional accuracy and dynamic deformation of 3D-printed surgical templates: an in vitro study.


ABSTRACT:

Purpose

To demonstrate the viability of a coordinate-measuring machine (CMM) for the geometric analysis of 3D printed surgical templates.

Methods

The template was designed and modified by adding 18 cylindrical landmarks for CMM test and then classified into five groups according to the slicing software and resins (opaque and transparent): Streamflow-O, Streamflow-T, Shapeware-T, Rayware-T and Polydevs-T (N = 3). Three standing times (0 w, 1 w, and 2 w) were included to observe possible deformation. All the measurements were performed automatically by the CMM through a preset program. The Euclidian distance (dxyz) was regarded as the representation of global dimension accuracy, and displacements in the x-, y-, and z-axes were also calculated.

Results

The average dxyz v