A multicenter evaluation of a novel microfluidic rapid AST assay for Gram-negative bloodstream infections.


ABSTRACT: Common phenotypic methods for antimicrobial susceptibility testing (AST) of bacteria are slow, labor intensive, and display considerable technical variability. The QuickMIC system provides rapid AST using a microfluidic linear gradient. Here, we evaluate the performance of QuickMIC at four different laboratories with regard to speed, precision, accuracy, and reproducibility in comparison to broth microdilution (BMD). Spiked (n = 411) and clinical blood cultures (n = 148) were tested with the QuickMIC Gram-negative panel and compared with BMD for the 12 on-panel antibiotics, and 10 defined strains were run at each site to measure reproducibility. Logistic and linear regression analysis was applied to explore factors affecting assay performance. The overall essential agreement