Determination of Calcium Isotopes Using Inductive Coupled Plasma-Tandem Mass Spectrometry with Ozone-Induced Oxidation.
Ontology highlight
ABSTRACT: In this study, we propose a new method for determination of calcium (Ca) isotopes using inductive coupled plasma-tandem mass spectrometry (ICP-MS/MS). The most abundant isotope of Ca, 40Ca, has a spectrum identical with that of argon (Ar), which is often used as the carrier gas in ICP-MS analysis. The detection capability for 40Ca is thus significantly reduced in the presence of 40Ar. To avoid interference from 40Ar during the 40Ca measurements, ozone (O3) was used as the reaction gas in the ICP-MS/MS analysis. The reaction of Ca+ with O3 in the collision cell preferentially produced CaO3+. In contrast, the reaction between Ar+ and O3 produced ArO+,