Prostate magnetic resonance imaging (MRI) in patients with hip implants-presetting a protocol using a phantom.


ABSTRACT:

Background

Metal structures are a source of artifacts that significantly complicate the interpretation of magnetic resonance imaging (MRI). The use of prostate MRI as a preliminary test in men with a suspicion on prostate cancer leads to an increased use of the test. The aim of this study was to solve a clinically significant problem: to ensure the reduction of artifacts from metal hip implants during prostate MRI. Another goal was to evaluate the impact of artifact reduction methods on quantitative measurements.

Methods

The prostate gland (PG) phantom model was a cylinder filled with an aqueous solution of polyvinylpyrrolidone at the concentrations of 40%, 30%, and 20% [central zone (CZ), peripheral zone (PZ), and "lesion", respectively]. Phantom MRI study was conducted on