Monoenergetic reconstructions and iodine density maps for visualization of coronary artery stents using 8-cm dual-layer detector spectral computed tomography: an <i>in vitro</i> phantom study.


ABSTRACT:

Background

The effectiveness of coronary computed tomography (CT) angiography in assessing stent restenosis is hindered by heavy metal artifacts. This study aimed to evaluate the image quality of monoenergetic reconstructions and iodine density map for coronary stent imaging using an 8-cm dual-layer detector spectral CT.

Methods

In this study, 8 stents with a diameter <3 mm (group A) and 10 with a diameter ≥3 mm (group B) were placed in plastic tubes filled with iodinated contrast media and scanned. The internal diameter of the prepared stents was then measured by intravascular ultrasound. The reconstructed images included iodine density maps, conventional images, and different energy levels. The visualization of the stent lumen and stent structure was subjectively assessed