Unknown

Dataset Information

Interference from anti-drug antibodies on the quantification of insulin: a comparison of an LC-MS/MS assay and immunoassays.


ABSTRACT: Aim: This study aims to compare the anti-drug antibody (ADA) interference in four pharmacokinetic (PK) assays across different platforms (AlphaLISA, Gyrolab, LC-MS/MS) and to devise a strategy for ADA interference mitigation to improve the accuracy of measured drug in total PK assays.Materials & methods: Spiked test samples, created to achieve different ADA concentrations in human serum also containing an insulin analogue, were analyzed alongside pooled clinical samples using four assays.Results & conclusion: Interference was observed in all platforms. A novel approach using the Gyrolab mixing CD, including acid dissociation in the PK assay, significantly reduced interference and thereby improved relative error from >99% to ≤20% yielding measurements well within the ac

SUBMITTER: Varga GM 

PROVIDER: S-EPMC11485705 | biostudies-literature | 2024 Sep

REPOSITORIES: biostudies-literature