Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes.


ABSTRACT:

Purpose

To evaluate the agreement of ocular biometry measured using a swept-source optical coherence (Casia 2) and a dual Scheimpflug (Galilei G6) tomography in keratoconus.

Methods

This retrospective study included 102 eyes from 102 keratoconus patient examined using both devices. Parameters compared included flat (Kf) and steep (Ks) keratometry, astigmatism of anterior, posterior, and total keratometry, central (CCT) and thinnest (TCT) corneal thickness. To assess the agreement, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used.

Results

Anterior and total Ks and Kf showed moderate or good agreement with 95% limits of agreement (LoA) range over 9.75 D. Posterior Ks and Kf were lower in the Galilei G6 (all p < 0.001).