Comparison of 3D Gradient-Echo Versus 2D Sequences for Assessing Shoulder Joint Image Quality in MRI.
Ontology highlight
ABSTRACT: Background: Three-dimensional gradient-echo (3D-GRE) sequences provide isotropic or nearly isotropic 3D images, leading to better visualization of smaller structures, compared to two-dimensional (2D) sequences. The aim of this study was to prospectively compare 2D and 3D-GRE sequences in terms of key imaging metrics, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), glenohumeral joint space, image quality, artifacts, and acquisition time in shoulder joint images, using 1.5-T MRI scanner. Methods: Thirty-five normal volunteers with no history of shoulder disorders prospectively underwent a shoulder MRI examination with conventional 2D sequences, including T 1- and T 2-weighted fast spin echo (T1/T2w