ABSTRACT: Purpose
We aim to verify predictions showing T2 relaxation rate of nanoparticle clusters and its dependence on spacing, size, geometry, and pulse sequence.Methods
We performed a laboratory validation study using nanopatterned arrays of iron oxide nanoparticles to precisely control cluster geometry and image diverse samples using a 4.7T MRI scanner with a T2 -weighted fast spin-echo multislice sequence. We applied denoising and normalization to regions of interest and estimated relative R2 for each relevant nanoparticle array or nanocluster array. We determined significance using an unpaired two-tailed t-test or one-way analysis of variance and performed curve fitting.Results
We measured a density-dependent T2 effect (p = 8