Impact of different lighting conditions on the tooth-shade selection using intra-oral scanners: An <i>in-vitro</i> study.


ABSTRACT:

Objectives

To evaluate the accuracy in tooth shade selection under various lighting conditions using different devices, including a color spectrophotometer (CSP) and two intra-oral scanners (IOS: IOS-1 and IOS-2).

Methods

Tooth shade measurements were performed on a standardized shade guide (16 shades, A1-D4) using CSP, IOS-1 and IOS-2. These evaluations were carried out under three lighting conditions: (1) DL - device light source only, (2) RL - DL plus room light, and (3) CL - RL plus dental-chair light. Each shade was measured ten times per lighting condition with each device by a single investigator. Shade detection accuracy was defined as the percentage of matches to the known shade. Logistic regression models were used for statistical analyses (α = 0.05).

Res