ABSTRACT: Background
Recent multicenter, multivendor MRI-based R2* vs. liver iron concentration (LIC) calibrations (i.e., MCMV calibrations) may facilitate broad clinical dissemination of R2*-based LIC quantification. However, these calibrations are based on a centralized offline R2* reconstruction, and their applicability with vendor-provided R2* maps is unclear.Purpose
To determine R2* ranges of agreement between the centralized and three MRI vendors' R2* reconstructions.Study type
Prospective.Subjects
Two hundred and seven subjects (mean age 37.6 ± 19.6 years; 117 male) with known or suspected iron overload from four academic medical centers.Field strength/sequence
Standardized multiecho spoiled gradient echo sequence at 1.5 T and 3.0 T for R2* mapping and