A Wearable Extracorporeal CO<sub>2</sub> Removal System with a Closed-Loop Feedback.
Ontology highlight
ABSTRACT: Extracorporeal Carbon Dioxide Removal (ECCO2R) systems support patients with severe respiratory failure. Concurrent ambulation and physical therapy improve patient outcomes, but these procedures are limited by the complexity and size of the extracorporeal systems and rapid changes in patient metabolism and the acid-base balance. Here, we present the first prototype of a wearable ECCO2R system capable of adjusting to a patient's changing metabolic needs. Exhaust gas CO2 (EGCO2) partial pressure is used as an analog for blood CO2 partial pressure (pCO2). Twin blowers modulate sweep gas through the AL to achieve a desired target EGCO2. The integrated system was tested in vitro for 24 h with water, under varying simul
SUBMITTER: Zhang A
PROVIDER: S-EPMC11505386 | biostudies-literature | 2024 Sep
REPOSITORIES: biostudies-literature
ACCESS DATA