Virtual Instruments for Peak-Overlapping Studies to Determine Low- and High-Concentration Components with Ion Chromatography: Potassium and Sodium.
Ontology highlight
ABSTRACT: We developed the LabVIEW-based virtual instruments (VIs) to bridge a gap in commercial software and to enable systematic peak-overlapping studies to recognise the concentration levels enabling reliable simultaneous determination of major and minor constituents in samples with wide concentration proportions. The VIs were applied to a case study of the ion chromatographic determination of potassium as minor and sodium as a major ion with an IonPac CS12A column and 50 μL injection loop. Two successive studies based on multilevel two-factorial response surface experimental designs, (1) a model peak-overlapping study based on single-ion injections, and (2) an accuracy and precision study, provided guidelines for real sample analyses. By adjusting sample dilutions so that the sodium mass concent
SUBMITTER: Gros N
PROVIDER: S-EPMC11510477 | biostudies-literature | 2024 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA