Field-Applicable Loop-Mediated Isothermal Amplification for the Detection of Seven Common Human Papillomavirus Subtypes.


ABSTRACT: Persistent HPV infection is a major risk factor for the subsequent development of cervical cancer. LAMP is simple and suitable for field detection in the resource-limited settings. In this study, hydroxy naphthol blue (HNB)-based visual LAMP and evagreen-based fluorescent LAMP coupled with a microfluidic chip (LAMP-chip) were established for the field detection of seven subtypes of HPV. The analytical sensitivity was 19-233 copies/reaction. The overall clinical sensitivity was 97.35% for visual LAMP and 98.23% for LAMP-chip. Both LAMP assays exhibited 100% specificity and were completed in less than 50 min. Additionally, both assays did not require complicated nucleic acid extraction and purification steps. A complete quality control monitoring system (including internal control, positive