Feasibility and performance of spin-echo EPI MR elastography at 3 Tesla for staging hepatic fibrosis in the presence of hepatic iron overload.


ABSTRACT:

Purpose

To assess the feasibility and performance of MR elastography (MRE) for quantifying liver fibrosis in patients with and without hepatic iron overload.

Methods

This retrospective single-center study analyzed 139 patients who underwent liver MRI at 3 Tesla including MRE (2D spin-echo EPI sequence) and R2* mapping for liver iron content (LIC) estimation. MRE feasibility and diagnostic performance between patients with normal and elevated LIC were compared.

Results

Patients with elevated LIC (21%) had significantly higher MRE failure rates (24.1% vs. 3.6%, p < 0.001) compared to patients with normal LIC (79%). For those with only insignificant to mild iron overload (LIC < 5.4 mg/g; 17%), MRE failure rate did not differ significantly from patients without iron overl