Ontology highlight
ABSTRACT: Introduction
Artificial cervical disc replacement (ACDR) is a widely accepted surgical procedure in the treatment of cervical radiculopathy and myelopathy. However, some research suggests that ACDR may redistribute more load onto the facet joints, potentially leading to postoperative axial pain in certain patients. Earlier studies have indicated that facet tropism is prevalent in the lower cervical spine and can significantly increase facet joint pressure. The present study aims to investigate the changes in the biomechanical environment of the cervical spine after ACDR using different prosthese when facet tropism is present.Methods
A C2-C7 cervical spine finite element model was created. Symmetrical, moderate asymmetrical (7 degrees tropism), and severe asymmetrical (14 de
SUBMITTER: Li J
PROVIDER: S-EPMC11524021 | biostudies-literature | 2024 Oct
REPOSITORIES: biostudies-literature