Unknown

Dataset Information

UV-C disinfection of ultrasound probes: Challenges of uneven irradiance on complex surfaces.


ABSTRACT: Medical devices that contact non-intact skin or mucous membranes are considered semi-critical devices and must undergo high-level disinfection (HLD) before use. Studies have identified several potential limitations of UV-C for HLD of semi-critical medical devices, including a lack of data demonstrating that UV-C irradiance can be uniformly applied to complex surfaces that contain grooves, notches and imperfections. This study focused on ultrasound probes as commonly used medical devices to show the distribution of irradiance on these surfaces. An endocavity bi-plane probe and curved array surface probe with typical surface topology were 3D scanned and modelled and an array of UV-C light-emitting diodes (LEDs) irradiating the probe surfaces was simulated (simulated wavelength: 275nm [peak],

SUBMITTER: Yeoh L 

PROVIDER: S-EPMC11524522 | biostudies-literature | 2024

REPOSITORIES: biostudies-literature