ABSTRACT: Objective
To determine the accuracy of the AViTA BP636 upper-arm home blood pressure (BP) monitor in general populations with small to extra-large arms according to the ANSI/AAMI/ISO Universal Standard (ISO 81060-2 : 2018/Amd1 : 2020).Methods
The AViTA BPM636 was utilized on 85 adult participants with upper-arm circumferences ranging from 15 to 52 cm, and four cuffs of different sizes (small: 15-22 cm, normal: 22-33 cm, large: 33-42 cm, and extra large: 42-52 cm) were used in this study. The recruitment fulfills the criteria of the ANSI/AAMI/ISO Universal Standard 81060-2 : 2018/Amd1 : 2020 in the general population using the same arm sequential BP measurement method.Results
For validation criterion 1, the mean ± SD of the differences between the test device and ref