Application of the adaptive Monte Carlo method for uncertainty evaluation in the determination of total testosterone in human serum by triple isotope dilution mass spectrometry.
Ontology highlight
ABSTRACT: The measurement uncertainty is a crucial quantitative parameter for assessing the reliability of the result. The study aimed to propose a new budget for uncertainty evaluation of a reference measurement procedure for the determination of total testosterone in human serum. The adaptive Monte Carlo method (aMCM) was used for the propagation of probability distributions assigned to various input quantities to determine the uncertainty of the testosterone concentration. The basic principles of the propagation and the statistical analysis were described based on the experimental results of the quality control serum sample. The analysis of the number of Monte Carlo trials was discussed. The procedure of validation of the GUM uncertainty framework using the aMCM was also provided. The number of M
SUBMITTER: Liu G
PROVIDER: S-EPMC11525383 | biostudies-literature | 2024 Aug
REPOSITORIES: biostudies-literature