Hemocompatibility Evaluation of a Novel Ambulatory Pulmonary Assist System Using a Lightweight Axial-Flow Pump.
Ontology highlight
ABSTRACT: The Pulmonary Assist System (PAS) is currently under development as a wearable respiratory assist system. In this study, the hemocompatibility of the PAS's axial-flow mechanical pump (AFP) was compared to other contemporary mechanical pumps in an acute ovine model. The PAS was attached to a normal sheep in a venovenous configuration using one of three pumps: 1) AFP, 2) ReliantHeart HeartAssist 5 (control), or 3) Abbott Pedimag (control) (n = 5 each). Each sheep was supported on the PAS for 12 hours with two L/minute of blood flow and four L/minute of sweep gas. Hemolysis, coagulation, inflammation, and platelet activation and loss were compared among the groups. In this study, the plasma-free hemoglobin (pfHb) was less than 10 mg/dl in all groups. The pfHb was significantly lower in the AF
SUBMITTER: Hong Y
PROVIDER: S-EPMC11527577 | biostudies-literature | 2024 Nov
REPOSITORIES: biostudies-literature
ACCESS DATA