Correlation between subretinal tissue plasminogen activator and air injection rates with pressure in a retina mimicking model.
Ontology highlight
ABSTRACT: By investigating the correlation between the injection rate and pressure of subretinal tissue plasminogen activator (tPA) and air using a standard Viscous Fluid Control (VFC) system with a 38-gauge cannula, we aimed to establish guidelines for stable injections. We fabricated a retina mimicking model (RMM) with 0.25% agarose solution and an aluminum plate, and substituted submacular hemorrhage (SMH) and tPA with blood-mimicking fluid (BMF) and balanced salt solution (BSS), respectively. The diameter of the pre-bleb mimicking SMH in RMM was 1.30 ± 0.16 cm, increasing to 1.98 ± 0.24 cm and 1.83 ± 0.22 cm after bleb propagation with BSS and air, respectively. BSS injection rates were 2.86 ± 0.04 µl/sec, 6.74 ± 0.48 µl/sec and 8.55 ± 0.16 µl/sec at 8, 12, and 16 psi, respectively. Air injectio
SUBMITTER: Kim NH
PROVIDER: S-EPMC11527981 | biostudies-literature | 2024 Oct
REPOSITORIES: biostudies-literature
ACCESS DATA