Oscillometric blood pressure measurements on smartphones using vibrometric force estimation.


ABSTRACT: This paper proposes a smartphone-based method for measuring Blood Pressure (BP) using the oscillometric method. For oscillometry, it is necessary to measure (1) the pressure applied to the artery and (2) the local blood volume change. This is accomplished by performing an oscillometric measurement at the finger's digital artery, whereby a user presses down on the phone's camera with steadily increasing force. The camera is used to capture the blood volume change using photoplethysmography. We devised a novel method for measuring the force applied of the finger without the use of specialized smartphone hardware with a technique called Vibrometric Force Estimation (VFE). The fundamental concept of VFE relies on a phenomenon where a vibrating object is dampened when an external force is appli