ABSTRACT: Purpose
To assess the accuracy of swept-source optical coherence tomography (SS-OCT) in detecting complete posterior vitreous detachment (PVD) in comparison with intraoperative findings.Methods
The retrospective study included 145 eyes of 145 consecutive patients who underwent surgery for epiretinal membranes or macular holes. Within a week prior to surgery, PVD status was evaluated by SS-OCT with a depth of field of 3 mm and a capture window of 16 × 8 mm. Complete PVD was identified when the hyaloid condensation was visible clearly on any B-scan or when the vitreous cortex reflectivity was not visible on all 33 B-scans. Sensitivity, specificity, positive predictive value, and negative predictive value of SS-OCT for detection of complete PVD were then compared with those ev