Safety and Feasibility Evaluation of Automated User Profile Settings Initialization and Adaptation With Control-IQ Technology.


ABSTRACT:

Background

Optimization of automated insulin delivery (AID) settings is required to achieve desirable glycemic outcomes. We evaluated safety and efficacy of a computerized system to initialize and adjust insulin delivery settings for the t:slim X2 insulin pump with Control-IQ technology in adults with type 1 diabetes (T1D).

Methods

After a 2-week continuous glucose monitoring (CGM) run-in period, adults with T1D using multiple daily injections (MDI) (N = 33, mean age 36.1 years, 57.6% female, diabetes duration 19.7 years) were transitioned to 13 weeks of Control-IQ technology usage. A computerized algorithm generated recommendations for initial pump settings (basal rate, insulin-to-carbohydrate ratio, and correction factor) and weekly follow-up settings to optimize glycemic