Comparison of the safety of flexible ureteroscopy with the different irrigation methods in a 3D print kidney model.


ABSTRACT:

Objective

To compare intrarenal pressure (IRP) and irrigation flow by varying suspended water heights and hand-held pressure pumping during flexible ureteroscopy using an in vitro 3D printed kidney model.

Methods

A 3D-printed silicone model was used to simulate the kidney. The ureteral access sheath(UAS) was connected to the kidney model and positioned at the ureteropelvic junction. Central venous pressure tubing was used to monitor the pressure in the renal pelvis under different conditions. Sheath sizes of 12Fr and 14Fr were tested with flexible ureteroscope (fURS) sizes of 7.5, 8.5, and 9.5Fr, respectively. The irrigation was gravity-based, with suspended water heights set at 60, 90, 120, 150, and 180 cm. The manual pumping as another set of measurement is used to measure

SUBMITTER: Bai B 

PROVIDER: S-EPMC11539779 | biostudies-literature | 2024 Nov