Single-Cell Lipidomics: An Automated and Accessible Microfluidic Workflow Validated by Capillary Sampling.
Ontology highlight
ABSTRACT: We report the first demonstration of a microfluidics-based approach to measure lipids in single living cells using widely available liquid chromatography mass spectrometry (LC-MS) instrumentation. The method enables the rapid sorting of live cells into liquid chambers formed on standard Petri dishes and their subsequent dispensing into vials for analysis using LC-MS. This approach facilitates automated sampling, data acquisition, and analysis and carries the additional advantage of chromatographic separation, aimed at reducing matrix effects present in shotgun lipidomics approaches. We demonstrate that our method detects comparable numbers of features at around 200 lipids in populations of single cells versus established live single-cell capillary sampling methods and with greater throughp
SUBMITTER: Kontiza A
PROVIDER: S-EPMC11541894 | biostudies-literature | 2024 Nov
REPOSITORIES: biostudies-literature