Unknown

Dataset Information

Analytical Sensitivity Analysis and Clinical Impact Modeling of Rapigen Rapid Diagnostic Tests for Malaria.


ABSTRACT: Laboratory benchmarking allows objective analysis of the analytical performance of malaria rapid diagnostic tests (RDTs). We present the analytical detection limits of the Rapigen BIOCREDIT Malaria Ag Pf/Pv (pLDH/pLDH), the Rapigen BIOCREDIT Malaria Ag Pf (pLDH/HRPII), and two best-in-class WHO-prequalified comparator RDTs, generated using standardized panels containing recombinant antigen, in vitro cultured parasites, international standards, and clinical samples. Detection limit antigen concentrations of HRP2, PfLDH, and PvLDH were determined for the Rapigen and comparator RDTs. Detection of antigens in international units (IU)/mL was also evaluated. The Rapigen Ag Pf (pLDH/HRPII) detected 3.9 and 3.9 IU/mL for PfLDH and HRP2, respectively, and the Ag Pf/Pv (pLDH/pLDH) detected 3.9 and 5

SUBMITTER: Golden A 

PROVIDER: S-EPMC11542527 | biostudies-literature | 2024 Nov

REPOSITORIES: biostudies-literature