Evaluation of the Accuracy of 3D Printed Patient-Specific Brain Biopsy Guide Using 3D Volume Rendering Technique in Canine Cadavers.
Ontology highlight
ABSTRACT: The objective of this study was to evaluate the accuracy of a CT-based, 3D-printed, patient-specific brain biopsy guide (3D-psBBG) through the application of a transfrontal approach in canine cadavers. A total of ten canine cadavers, with weights ranging from 4.36 to 14.4 kg, were subjected to preoperative CT scans to generate 3D skull models. Customized biopsy guides were created based on these models and manufactured using 3D printing technology. Twenty spinal needle insertions were performed, and the accuracy of needle placement was evaluated through both CT and 3D volume-rendering techniques. The mean needle placement error was 2.1 mm, with no significant differences observed between insertions targeting the fronto-olfactory and piriform lobes. The 3D volume-rendering method demonstrat
SUBMITTER: Jeong M
PROVIDER: S-EPMC11545512 | biostudies-literature | 2024 Nov
REPOSITORIES: biostudies-literature
ACCESS DATA