Unknown

Dataset Information

Development of a digital amplifier system for cut-open oocyte electrophysiology.


ABSTRACT: The cut-open oocyte Vaseline gap technique is a powerful electrophysiological method for the characterization of ion channels. However, traditional amplifiers for cut-open oocyte Vaseline gap are labor intensive and require significant user expertise. We introduce an innovative, open-source digital amplifier system with high-speed digitization and software-controlled electronics for computer-driven automation. This system compares well to existing commercial systems in terms of conventional specifications of step response (current peak at 25μs and decay of 36μs time constant), current noise (1.0 nA at 3-kHz bandwidth), and dynamic range (96.9 dB). Additionally, it unlocks new methods through close integration of the amplifier and software, including machine-learning techniques for tuning c

SUBMITTER: Koerner LJ 

PROVIDER: S-EPMC11549981 | biostudies-literature | 2024 Oct

REPOSITORIES: biostudies-literature