Comparison of Hemodynamic Performance, Three-Dimensional Flow Fields, and Turbulence Levels for Three Different Heart Valves at Three Different Hemodynamic Conditions.


ABSTRACT: The hemodynamic performance of different prosthetic heart valves is difficult to compare among studies due to a variety of test conditions and experimental techniques. Existing studies are typically limited to one family of valves (biological or mechanical) and testing conditions of 5l/min and often lack sufficient spatial resolution. To address these limitations, a pulse duplicator with a multi-view imaging system (Tomo-PIV) was employed to investigate the three-dimensional flow field in the aortic root of three different valves: a tri-leaflet mechanical heart valve (TRIFLO, Novostia), a bi-leaflet mechanical heart valve (On-X, Artivion), and a biological heart valve (Perimount, Edwards Lifesciences). The valves were tested at low (3 l/min), normal (5 l/min), and elevated (7 l/min) cardia