Remote Sensing of Nitric Acid and Temperature via Design of Experiments, Chemometrics, and Raman Spectroscopy.
Ontology highlight
ABSTRACT: This study presents an effective method for the quantification of nitric acid (0.1-9 M) and the temperature (20-60 °C) through optimal experimental design, chemometrics, and Raman spectroscopy. Raman spectroscopy can be deployed using fiber-optic cables in hot cell environments to support processing operations in the nuclear field and industry. Chemical operations frequently use nitric acid and operate at nonambient temperatures either by design or by circumstance. Examples of Raman spectroscopy for the quantification of nitric acid with applications in the industrial field are profuse. However, the effect of temperature on quantification is often ignored and should be considered in real-world scenarios. Statistical design of experiments was used to build training sets for partial least-sq
SUBMITTER: Russell DV
PROVIDER: S-EPMC11561608 | biostudies-literature | 2024 Nov
REPOSITORIES: biostudies-literature