ABSTRACT:

Background

To investigate the artifact sizes of four common breast clip-markers on a standard breast magnetic resonance imaging (MRI) protocol in an in vitro phantom model.

Methods

Using 1.5-T and 3-T whole-body scanners with an 18-channel breast coil, artifact dimensions of four breast biopsy markers in an agarose-gel phantom were measured by two readers on images obtained with the following sequences: T2-weighted fast spin-echo short inversion time fat-suppressed inversion-recovery with magnitude reconstruction (T2-TIRM); T1-weighted spoiled gradient-echo with fat suppression (T1_FL3D), routinely used for dynamic contrast-enhanced imaging; diffusion-weighted imaging (DWI), including a readout segmented echo-planar imaging (RESOLVE-DWI) and echo-planar imaging sequence (EPI