ABSTRACT: Background and purpose
Dosimetry audits are essential for reducing errors in brachytherapy. A postal dosimetry audit methodology was developed and tested in an international multicentre pilot, to assess the accuracy of the Reference Air Kerma Rate of 192Ir and 60Co brachytherapy sources.Materials and methods
A compact phantom made of polymethyl methacrylate was developed to accommodate two catheters, a radiophotoluminescence dosimeter (RPLD) for dose measurements and a Gafchromic (RTQA2) film strip for source position verification. Deviations of the audit setup from TG-43 conditions were quantified experimentally and compared to previous Monte Carlo (MC) simulations. A measurement uncertainty budget was estimated for the RPLD analysis. The methodology