Accuracy and Feasibility of a Novel Glucose/Lactate Continuous Multi-Analyte Sensing Platform in Humans.


ABSTRACT:

Background and aim

Continuous glucose monitoring systems (CGMs) have been commercially available since 1999. However, automated insulin delivery systems may benefit from real-time inputs in addition to glucose. Continuous multi-analyte sensing platforms will meet this area of potential growth without increasing the burden of additional devices. We aimed to generate pilot data regarding the safety and function of a first-in-human, single-probe glucose/lactate multi-analyte continuous sensor.

Methods

The investigational glucose/lactate continuous multi-analyte sensor (PercuSense Inc, Valencia, California) was inserted to the upper arms of 16 adults with diabetes, and data were available for analysis from 11 of these participants (seven female; mean [SD] = age 43 years [16]; bo