Analytical measuring interval, linearity, and precision of serology assays for detection of SARS-CoV-2 antibodies according to CLSI guidelines.
Ontology highlight
ABSTRACT: Serology testing is commonly used to evaluate the immunogenicity of COVID-19 vaccines and measure antibodies as a marker of previous infection with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). In this study, four laboratory-developed serology enzyme-linked immunosorbent assays (SARS-CoV-2 anti-Spike and anti-Nucleocapsid immunoglobin G [IgG] and immunoglobin M [IgM]) calibrated to the WHO International Standard 20/136 were validated via analytical measuring interval (limit of blank [LOB], limit of detection [LOD], and limit of quantification [LOQ]), linearity, and precision according to the Clinical and Laboratory Standards Institute (CLSI) guidelines EP17-A2, EP06 2nd Edition, and EP05-A3. For Spike IgG, LOB was 3.0 binding antibody units per milliliter (BAU/mL), LOD was
SUBMITTER: Haynesworth K
PROVIDER: S-EPMC11580426 | biostudies-literature | 2024 Nov
REPOSITORIES: biostudies-literature