Long-Term Accurate Continuous Glucose Biosensors via Extended Nitric Oxide Release.
Ontology highlight
ABSTRACT: Analytical performance and tissue interactions of nitric oxide (NO)-releasing continuous glucose sensors were evaluated over a 28 d study in a diabetic swine model. Interstitial glucose was detected using an implanted needle-type amperometric glucose sensor. Two NO-release durations from the sensor surface were achieved by doping the membranes with nonporous (14 d release) or porous (30 d release) S-nitrosothiol-functionalized silica nanoparticles. Numerical and clinical accuracy of the sensors were assessed at time points (1, 7, 14, 21, and 28 d) following implantation. Nitric oxide-releasing sensors demonstrated accurate glucose detection over a time period directly correlated with the active release of NO. Silica particle-doped sensors that released NO for 30 d showed standard-co
SUBMITTER: Malone-Povolny MJ
PROVIDER: S-EPMC11581131 | biostudies-literature | 2019 Dec
REPOSITORIES: biostudies-literature
ACCESS DATA