Reliability and minimal detectable change of the MX3 hydration testing system.


ABSTRACT: Assessing hydration status outside of laboratories can be challenging given that most hydration measures are invasive, stationary, costly, or have questionable validity. This study investigated the within-day, test-retest reliability, and minimal detectable change (MDC) of the MX3 Hydration Testing System (MX3 Diagnostics), a relatively low cost, noninvasive, and portable method to measure saliva osmolality. Seventy-five adults (44 men, 31 women; 29.6±10.8 yr, 171.1±9.2 cm, 79.1±15.4 kg) presented two saliva samples approximately 3 to 5 minutes apart. Fluid intake was avoided for at least 5 minutes prior to sample collections. For each sample collection, a researcher used the MX3 to tap the tip of a test strip to saliva on a participant's tongue. Intra-class correlation coefficients (ICCs)