Design consideration on integration of mechanical intravascular ultrasound and electromagnetic tracking sensor for intravascular reconstruction.


ABSTRACT:

Purpose

Considering vessel deformation, endovascular navigation requires intraoperative geometric information. Mechanical intravascular ultrasound (IVUS) with an electromagnetic (EM) sensor can be used to reconstruct blood vessels with thin diameter. However, the integration design should be evaluated based on the factors affecting the reconstruction error.

Methods

The interference between the mechanical IVUS and EM sensor was measured in different relative positions. Two designs of the integrated catheter were evaluated by measuring the reconstruction errors using a rigid vascular phantom.

Results

When the distance from the EM sensor to the field generator was 75 mm, the interference from mechanical IVUS to an EM sensor was negligible, with position and rotation erro