Verification of the Vitek Reveal System for Direct Antimicrobial Susceptibility Testing in Gram-Negative Positive Blood Cultures.


ABSTRACT: Background/Objectives: The International Organization for Standardization (ISO) 20776-2:2021, which replaces ISO 20776-2:2007, focuses solely on the performance of antimicrobial susceptibility testing (AST) assays, emphasizing the ISO 20776-1 broth microdilution method as the reference standard. Consequently, categorical agreement (CA) and associated errors should not be applied. We verified the Vitek Reveal AST assay according to both ISO 20776-2:2021 and ISO 20776-2:2007 criteria. Methods: Samples from 100 simulated and clinical Gram-negative (GN) positive blood cultures (PBCs) were tested at a large teaching hospital. The simulated GN-PBCs were obtained from a hospital collection of isolates selected to represent diverse antimicrobial resistance profiles. The Reveal assay