Unknown

Dataset Information

Evolution of a laboratory mechanomyograph.


ABSTRACT: Mechanomyography is currently the accepted laboratory reference standard for quantitative neuromuscular blockade monitoring. Mechanomyographs are not commercially available. Previously, a mechanomyograph was built by our laboratory and used in several clinical studies. It was subsequently redesigned to improve its usability and functionality and to accommodate a wider range of hand sizes and shapes using an iterative design process. Each version of the redesigned device was initially tested for usability and functionality in the lab with the investigators as subjects without electrical stimulation. The redesigned devices were then assessed on patients undergoing elective surgery under general anesthesia without neuromuscular blocking drugs. Since the patients were not paralyzed, the expect

SUBMITTER: Wedemeyer Z 

PROVIDER: S-EPMC11604740 | biostudies-literature | 2024 Dec

REPOSITORIES: biostudies-literature