Ontology highlight
ABSTRACT: Purpose
Phosphorus MRS (31P MRS) enables noninvasive assessment of energy metabolism, yet its application is hindered by sensitivity limitations. To overcome this, often high magnetic fields are used, leading to challenges such as spatial B 1 + $$ {\mathrm{B}}_1^{+} $$ inhomogeneity and therefore the need for accurate flip-angle determination in accelerated acquisitions with short TRs. In response to these challenges, we propose a novel short TR and look-up table-based double-angle method for fast 3D 31P B 1 + $$ {\mathrm{B}}_1^{+} $$ mapping (fDAM).Methods
Our method incorporates 3D weighted stack-of-spiral gradient-echo acquisitions and a frequency-selective pulse to enable efficient B 1 + $$ {\mathrm{B}}_1^{+} $$ mapping based on the pho
SUBMITTER: Widmaier MS
PROVIDER: S-EPMC11604843 | biostudies-literature | 2025 Feb
REPOSITORIES: biostudies-literature