ABSTRACT: Purpose
Transmit arrays for body imaging have characteristics of both volume and local transmit coils. This study evaluates two specific absorption rate (SAR) aspects, local and whole-body SAR, of arrays for body imaging at 7 T and also for a 3 T birdcage.Methods
Simulations were performed for six antenna arrays at 7 T and one 3 T birdcage. Local SAR matrices and the whole-body SAR matrix were computed and evaluated with random shims. A set of reduced local SAR matrices was determined by removing all matrices dominated by the whole-body SAR matrix.Results
The results indicate that all RF transmit coils for body imaging in this study are constrained by the local SAR limit. The ratio between local and whole-body SAR is nevertheless smaller for arrays with large FOV, a