Impact of Leg Position on Measurements Used to Detect Femoral Component Subsidence in THA.
Ontology highlight
ABSTRACT: A fully automated artificial intelligence-based tool was developed to detect and quantify femoral component subsidence between serial radiographs. However, it did not account for measurement errors due to leg position differences, such as rotation or flexion, between comparative radiographs. If there are small differences in rotation or flexion of the leg between comparative radiographs, the impact on subsidence measurement is unclear. Twenty-five primary total hip arthroplasty procedures were performed by 3 fellowship-trained arthroplasty surgeons using a direct anterior approach. A Hana table allowed precise changes in femur position. Final fluoroscopic images were collected with rotational and flexion changes applied to the femur without moving the C-arm. Subsidence values were manually