Feasibility of Assessment of Habitual Salt Intake Using a 24-h Urinary Salt Excretion Self-Measuring Device.


ABSTRACT:

Introduction

To assess habitual salt intake, tools are needed to measure 24-h urinary salt excretion repeatedly. We developed and validated a new portable salt monitor, which measures salt excreted per urination and sums the values to provide an accurate estimate of urinary salt excretion over 24 h.

Methods

A previously developed salt monitor was improved with respect to the capacity, volume sensors, and equation for urinary sodium chloride concentration estimation. In 20 healthy Japanese female volunteers, 24-h urinary salt excretion was measured using the salt monitor and a conventional 24-h urine collection method on eight nonconsecutive days.

Results

In a total of 157 days, there were no fixed or proportional errors between the methods. The mean salt intake over 8

SUBMITTER: Morikawa N 

PROVIDER: S-EPMC11614415 | biostudies-literature | 2024