Fat-free noncontrast whole-heart cardiovascular magnetic resonance imaging with fast and power-optimized off-resonant water-excitation pulses.


ABSTRACT:

Background

Cardiovascular magnetic resonance imaging (CMR) faces challenges due to the interference of bright fat signals in visualizing structures, such as coronary arteries. Effective fat suppression is crucial, especially when using whole-heart CMR techniques. Conventional methods often fall short due to rapid fat signal recovery, leading to residual fat content hindering visualization. Water-selective off-resonant radiofrequency (RF) pulses have been proposed but come with tradeoffs between pulse duration, which increases scan time, and increased RF energy deposit, which limits their applicability due to specific absorption rate (SAR) constraints. The study introduces a lipid-insensitive binomial off-resonant (LIBOR) RF pulse, which addresses concerns about SAR and scan time, a

SUBMITTER: Mackowiak ALC 

PROVIDER: S-EPMC11616052 | biostudies-literature | 2024 Winter