ABSTRACT: Aims
The aim was to evaluate the effect of extended use of the Omnipod® 5 Automated Insulin Delivery (AID) System in adults with type 2 diabetes and suboptimal glycaemic control.Materials and methods
Following an 8-week single-arm, multicentre, outpatient trial of AID in adults with type 2 diabetes and baseline HbA1c ≥ 8% (≥ 64 mmol/mol), participants were given the opportunity to continue use of the AID system in a 26-week (~6 month) extension phase. The primary safety endpoints were percentage of time with sensor glucose ≥ 250 mg/dL and < 54 mg/dL. Additional glycaemic measures, including percentage of time in range (TIR) (70-180 mg/dL) and HbA1c, were evaluated. The use of non-insulin anti-hyperglycaemic medications was permitted throughout the entire study.Results</