Unknown

Dataset Information

CNN-Based Device-Agnostic Feature Extraction From ONH OCT Scans.


ABSTRACT:

Purpose

Optical coherence tomography (OCT)-derived measurements of the optic nerve head (ONH) from different devices are not interchangeable. This poses challenges to patient follow-up and collaborative studies. Here, we present a device-agnostic method for the extraction of OCT biomarkers using artificial intelligence.

Methods

ONH-centered OCT volumes from the Heidelberg SPECTRALIS, ZEISS CIRRUS HD-OCT 5000, and Topcon 3D OCT-1000 Mark I/II and 3D OCT-2000 devices were annotated by trained graders. A convolutional neural network (CNN) was trained on these segmented B-scans and utilized to obtain several ONH biomarkers, such as the retinal nerve fiber layer (RNFL) and the minimal rim width (MRW). The CNN results were compared between different devices and to the manufacturer

SUBMITTER: Driessen SJ 

PROVIDER: S-EPMC11620008 | biostudies-literature | 2024 Dec

REPOSITORIES: biostudies-literature