Ontology highlight
ABSTRACT: Objective
To describe a novel Small Fragment Removal System (SFRS) which is hypothesized to improve stone fragment removal during flexible ureteroscopy in patients with urolithiasis. The SFRS consists of three parts: a Syphon Ureteric Access Sheath (SUAS), a Dual Action Pump (DAP) and an Agitator. This bench assessment aims to assess the SFRS's impact on intra-renal pressure (IRP), irrigant flow rate and stone fragment removal compared to a traditional UAS.Materials and methods
A validated phantom kidney and fibre optic pressure sensor was used to assess IRP. Standardized irrigation via a flexible ureterorenoscope was instilled through a traditional UAS and compared to the novel SFRS. Both were 11/13Fr in size. Measured minute volume, IRP and percentage of stone fragments r
SUBMITTER: Lazarus J
PROVIDER: S-EPMC11624219 | biostudies-literature | 2024 Dec